Technical Specifications
● System Function
Automatic, Random Access
STAT sample priority, available for all sample position
Throughput: Constant throughput 60 tests/hour (double reagent); 90 tests/hour (single
reagent)
Insert PC system with touch screen
Contents of the testing: Clinical chemistry, Immunoturbidimetric
Analytical modes: End-point, Two-point, Kinetic,
● System Tray
Sample position: 9 positions for samples
Sample volume: 3~ 45μL, 0.5μL adjustable
Flexible sample position design: MAX: 42 sample positions
● Reagent Tray
Reagent position: 26 positions for reagent
Reagent volume:
R1: 180~450μL, 1μL adjustable
R2: 3~450μL, 1μL adjustable
● Reaction System:
Reaction positions: 90 cuvettes, 10 cuvettes/strip
Cuvette: Optical length 5mm
Reaction volume: 180 ~ 500ul
Reaction temperature: 37℃, fluctuation ±0.3℃
1 mixing probe
● Probe
Pro-heating for reagent
Collision protection, Liquid level detection and Inventory checking
Automatic inner and outer probe washing after every cycle, carry-over<0.1%
● Measuring and Optic System
Lamp: Halogen Lamp
Absorbance range: 0 ~ 3.500 ABS
Wavelengths: 8 wavelengths from 320 ~ 850 nm
Resolution: 0.0001 ABS
● Working Condition
Power supply: 100 ~ 240 VAc 50/60 Hz,1500w
Temperature: 10 ~ 35°C
Humidity: 35 ~ 80%
Water consumption: <3 L/hour
Dimention: 630mm(L)480mm(W)510mm(H)
Net Weight: 70 KG